Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Simpson 260-3: (1 year) CAL VER /5500 DATE: 09-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 33 NUMBER OF LINES: 157 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: None 1.004 DISP Temperature: 0 - 50degC 1.004 DISP Humidity: <70% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Throughout this procedure, make appropriate 1.006 DISP connection changes for ranges with seperate inputs. 1.007 HEAD {MECHANICAL METER ZERO} 1.008 DISP Place the UUT in a horizontal position, and if necessary 1.008 DISP adjust the FRONT PANEL mechanical meter zero for a ZERO 1.008 DISP indication. Be sure to relieve the tension on the zero 1.008 DISP control by turning in the opposite direction slightly. 1.009 RSLT =If necessary, FRONT PANEL meter zero was adjusted. 1.010 HEAD {DC VOLTAGE} 1.011 DISP Connect the 5500A to the UUT as follows: 1.011 DISP [32] 5500A NORMAL HI to UUT POS 1.011 DISP [32] 5500A NORMAL LO to UUT COMMON 1.012 HEAD DC VOLTAGE: {50V Range} 1.013 5500 50 50.0V 1.5U 2W 2.001 5500 50 40.0V 1.5U 2W 3.001 5500 50 30.0V 1.5U 2W 4.001 5500 50 20.0V 1.5U 2W 5.001 5500 50 10.0V 1.5U 2W 6.001 HEAD DC VOLTAGE: {2.5V Range} 6.002 5500 2.5 2.500V 0.075U 2W 7.001 HEAD DC VOLTAGE: {10V Range} 7.002 5500 10 10.0V 0.3U 2W 8.001 HEAD DC VOLTAGE: {250V Range} 8.002 5500 250 250.0V 7.5U 2W 9.001 HEAD DC VOLTAGE: {1000V Range} 9.002 5500 1000 1000V 30U 2W 10.001 DISP Connect the 5500A to the UUT as follows: 10.001 DISP [32] 5500A NORMAL HI to UUT 5000VDC 10.001 DISP [32] 5500A NORMAL LO to UUT COMMON 10.002 ASK- R 10.003 HEAD DC VOLTAGE: {5000V Range} 10.004 5500 5000 1000V 150U 2W 11.001 ASK+ R 11.002 HEAD {AC VOLTAGE} 11.003 DISP Connect the 5500A to the UUT as follows: 11.003 DISP [32] 5500A NORMAL HI to UUT POS 11.003 DISP [32] 5500A NORMAL LO to UUT COMMON 11.004 HEAD AC VOLTAGE: {2.5V Range} 11.005 5500 2.5 2.500V 0.075U 60H SI 2W 12.001 5500 2.5 2.000V 0.075U 60H SI 2W 13.001 5500 2.5 1.500V 0.075U 60H SI 2W 14.001 5500 2.5 0.500V 0.075U 60H SI 2W 15.001 HEAD AC VOLTAGE: {50V Range} 15.002 5500 50 50.0V 1.5U 400H SI 2W 16.001 5500 50 40.0V 1.5U 400H SI 2W 17.001 5500 50 30.0V 1.5U 400H SI 2W 18.001 5500 50 20.0V 1.5U 400H SI 2W 19.001 5500 50 10.0V 1.5U 400H SI 2W 20.001 HEAD AC VOLTAGE: {10V Range} 20.002 5500 10 10.0V 0.3U 60H SI 2W 21.001 HEAD AC VOLTAGE: {250V Range} 21.002 5500 250 250.0V 7.5U 60H SI 2W 22.001 HEAD AC VOLTAGE: {1000V Range} 22.002 5500 1000 1000V 30U 60H SI 2W 23.001 DISP Connect the 5500A to the UUT as follows: 23.001 DISP [32] 5500A NORMAL HI to UUT 5000VAC 23.001 DISP [32] 5500A NORMAL LO to UUT COMMON 23.002 HEAD AC VOLTAGE: {5000V Range} 23.003 ASK- R 23.004 5500 5000 1000V 150U 60H SI 2W 24.001 ASK+ R 24.002 HEAD {CURRENT} 24.003 DISP Connect the 5500A to the UUT as follows: 24.003 DISP [32] 5500A AUX HI to UUT 50uA 24.003 DISP [32] 5500A AUX LO to UUT COMMON 24.004 HEAD CURRENT: {50uA Range} 24.005 5500 50 50.0uA 1.5U 2W 25.001 DISP Connect the 5500 to the UUT as follows: 25.001 DISP [32] 5500A AUX HI to UUT POS 25.001 DISP [32] 5500A AUX LO to UUT COMMON 25.002 HEAD CURRENT: {1mA Range} 25.003 5500 1 1.00mA 0.03U 2W 26.001 HEAD CURRENT: {10mA Range} 26.002 5500 10 10.0mA 0.3U 2W 27.001 HEAD CURRENT: {100mA Range} 27.002 5500 100 100mA 3U 2W 28.001 HEAD CURRENT: {500mA Range} 28.002 5500 500 500mA 15U 2W 29.001 DISP Connect the 5500 to the UUT as follows: 29.001 DISP [32] 5500A AUX HI to UUT +10 AMPS 29.001 DISP [32] 5500A AUX LO to UUT -10 AMPS 29.002 HEAD CURRENT: {10A Range} 29.003 5500 10 5.0A 0.3U 2W 30.001 HEAD {RESISTANCE} 30.002 ASK- R 30.003 DISP Connect the 5500A to the UUT as follows: 30.003 DISP [32] 5500A AUX HI to UUT V/OHM 30.003 DISP [32] 5500A AUX LO to UUT COM 30.003 DISP [32] 5500A NORMAL HI to UUT V/OHM 30.003 DISP [32] 5500A NORMAL LO to UUT COM 30.003 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 30.003 DISP the NORMAL leads to the SENSE leads AT the UUT. 30.004 DISP Set the Function Switch to +DC 30.005 HEAD RESISTANCE: {Rx1 Range} 30.006 DISP Set the Range Switch to Rx1 30.007 5500 0.0Z S CW 30.008 DISP Adjust the front panel ZERO OHMS control for a meter 30.008 DISP indication of 0 Ohms. 30.008 DISP 30.008 DISP If unit will not zero, replace the D-sized battery. 30.009 5500 10.0Z 1U CW 31.001 HEAD RESISTANCE: {Rx100 Range} 31.002 DISP Set the Range Switch to Rx100 31.003 5500 0.0Z S CW 31.004 DISP Adjust the front panel ZERO OHMS control for a meter 31.004 DISP indication of 0 Ohms. 31.004 DISP 31.004 DISP If unit will not zero, replace the D-sized battery. 31.005 5500 1000Z 100U CW 32.001 HEAD RESISTANCE: {Rx10,000 Range} 32.002 DISP Set the Range Switch to Rx10,000 32.003 5500 0.0Z S CW 32.004 DISP Adjust the front panel ZERO OHMS control for a meter 32.004 DISP indication of 0 Ohms. 32.004 DISP 32.004 DISP If unit will not zero, replace the D-sized battery. 32.005 5500 100kZ 10U CW 33.001 END